Tracheostomy Tubes, Air Lon Inhalation Set (Inner Cannula with 15 mm Adapter) Nylon, item number 1050180 size 3, item number 1050182 size 4, item number 1050184 size 5, item number 1050186 size 6, item number 1050188 size 7, and item number 1050190 size 8.

FDA Recall Z-1466-04 · Posted 2004-09-15

Recall StatusTerminated
Product CodeJOH
Date Initiated2004-08-05
Date Terminated2005-11-23
Product Quantity22,324 tubes

Reason for Recall

foreign objects (polishing stones) in the tubes

Firm Action

The recalling firm issued a recall letter dated 8/5/04 to their distributors. The letter informs the accounts of the problem and the need to return any affected product. The recalling firm issued a second letter dated 8/11/04 to the distributors with instructions to conduct a sub-recall.

Root Cause

Other

Distribution Pattern

The products were shipped to 45 distributors nationwide. The products were also shipped to Canada, Malaysia, Taiwan, Thailand, Turkey, and the Dominican Republic.