Terminated

Edwards Lifesciences Fem-Flex II Arterial and femoral Cannulae, arterial cannula sizes 16, 18 and 20 French, models: FEMII016A, FEMII016AS, FEMII018A, FEMII018AS, FEMII020A, and FEMII020AS.

FDA Recall Z-1466-2014 · Posted 2014-04-16

Recall StatusTerminated
Product CodeDWF
Date Initiated2014-04-09
Date Terminated2016-12-28
Product Quantity13,115

Reason for Recall

Potential for separation of dilator tips.

Firm Action

Customers were notified via letter on 4/4/14 to return affected products.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide, including foreign, govt/VA/Military consignees.