Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.

FDA Recall Z-1467-2025 · Posted 2025-03-26

Recall StatusOpen, Classified
Product CodeETN
Date Initiated2025-02-24
Product Quantity452

Reason for Recall

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Firm Action

The recalling firm alerted affected hospitals by visit on February 24, 2025. Customers were asked to check inventory for affected lot numbers and to return the affected inventory for replacement. Notifications were issued on February 26, 2025.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution.