Terminated

CODMAN CERTAS Programmable Valves, central nervous system shunt, models as follows: 82-8800 In Line Valve Only 82-8801 In Line Valve with Catheter and Accessories 82-8802 In Line Valve with Unitized Catheter and Accessories 82-8803 In line Valve with Unitized BACTISEAL Catheter and Accessories 82-8804 In Line Valve only with SIPHONGUARD Device 82-8805 In Line Valve with SIPHONGUARD Device, Catheter and Accessories 82-8806 In Line Valve with SIPHONGUARD, Unitized Catheter and Accessories 82-8807 In Line Valve with SIPHONGUARD Device, Unitized BACTISEAL Catheter and Accessories 82-8850 Certas Therapeutic Management System

FDA Recall Z-1468-2014 · Posted 2014-04-17

Recall StatusTerminated
Product CodeJXG
Date Initiated2013-12-13
Date Terminated2016-04-21
Product Quantity9,445 units

Reason for Recall

Ensuring all customers have the tools to assist in verifying the setting of the valve on x-ray images and the location and orientation of implanted CODMAN CERTAS Programmable Valves.

Firm Action

Codman Neuro issued notification packages dated 12/13/13 to Domestic Hospitals and international accounts. Specific letters and associated inclusion of the additional aids, include A) a valve setting verification guide with instructions for use with x-ray images, and B) a plastic valve cut out overlay). A response reply form was included to be completed and returned to confirm effectiveness of the notification.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →