HammerTube Implant, 2.75mm, 10 Cannulated (Sterile)
FDA Recall Z-1468-2022 · Posted None
Recalling FirmParagon 28, Inc.
Recall StatusOpen, Classified
Product CodeHTY
Date Initiated2022-05-26
Product Quantity9 implants
Reason for Recall
Due to Titanium plasma coating not present implant.
Firm Action
On 05/26/2022, the firm communicated via email to its customers/distributors informing them that Paragon 28 has identified via customer complaints that the titanium plasma coating is missing on the HammerTube Implants.
On 06/15/2022, Paragon28 communicated an updated "URGENT Medical Device Recall" letter to customer that informed them of the potential health risks involved with the affected devices such as:
" The defect may cause the implant to potentially result in reduced of pull-out force creating instability at the joint which may lead to delayed or lack of healing.
" The implants, with the defect, may cause potential mating interaction failure between inserter and implant.
" The defect may lead to an incorrect intended implant placement; directional and or depth.
Customers are instructed to:
1. Review physical inventory, discontinue use, quarantine and not to further distribute the affected devices.
2. Return affected devices immediately to:
Paragon28, Attn: Recalls
14445 Grasslands Dr.
Englewood, CO 80112
Any questions, contact:
Mandy Kinnee
Email: [email protected]
Phone: 720.966.8333
*Note: You may also contact [email protected].
1.
Root Cause
Under Investigation by firm
Distribution Pattern
U.S. Nationwide distribution in the states of AL, CA, GA, IL and KY.