Terminated

Zimmer Trabecular metal primary hip prosthesis femoral stem, press-fit, collarless, 12/14 neck taper - standard body -extended neck offset, size 14, 149 mm stem length, sterile, tivanium TI-6AL-4V alloy/tantalum, sterile; Cat. no. 00-7864-014-20 (00786401420).

FDA Recall Z-1471-06 · Posted 2006-08-31

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2006-08-02
Date Terminated2007-05-04
Product Quantity181 of all devices

Reason for Recall

Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly.

Firm Action

Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland.