Campy CVA Medium 100/PK, R01272
FDA Recall Z-1474-2026 · Posted 2026-02-26
Advertisement
Recalling FirmRemel, Inc
Recall StatusOpen, Classified
Product CodeJSJ
Date Initiated2026-02-05
Product Quantity97
Reason for Recall
Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium.
Firm Action
On February 5, 2026, the firm notified affected customers via Urgent Medical Device Recall letter. Customers were instructed to review test results and destroy any remaining inventory of the affected lot.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of AZ, CO, IL, NJ, OH, TN, WA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.