Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Zimmer ZMR Hip System femoral body, revision, nitrided, porous, 12/14 neck taper, standard, 40 mm neck offset, cone body, size C, 35 mm build-up, tivanium TI-6AL-4V alloy, sterile; REF. 9996-19-35 (Cat. no. 00999601935).

FDA Recall Z-1478-06 · Posted 2006-08-31

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2006-08-02
Date Terminated2007-05-04
Product Quantity181 of all devices

Reason for Recall

Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly.

Firm Action

Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑