NaviNetics Skull Anchor Key and Drill Kit, REF NN1215
FDA Recall Z-1478-2024 · Posted 2024-04-04
Advertisement
Recalling FirmNavinetics Inc
Recall StatusOpen, Classified
Product CodeHAW
Date Initiated2024-02-26
Product Quantity20 kits
Reason for Recall
Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.
Firm Action
NaviNetics issued an URGENT: NN1215 NAVINETICS SKULL ANCHOR KEY AND DRILL KIT RECALL notice to its consignees on 02/27/2024 via email. The notice explained the problem with the device, risk, and requested the return of the product.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of FL, MN, CO.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.