Luminos Agile Max (VE10, VF10, VF11)
FDA Recall Z-1479-2023 · Posted 2023-05-23
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2022-11-23
Product Quantity704 units in the United States (1945 units worldwide)
Reason for Recall
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube. Siemens Healthineers became aware of one customer site where this issue was identified during clinical operation.
Firm Action
Siemens Medical Solutions USA, Inc., will provide software updates to the
affected systems with software versions VF10 and VF11 with software via Update
Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide - Worldwide Distribution