Captiva's lumbar intervertebral fusion system Reamers (color code rings on the Reamers).
FDA Recall Z-1480-2015 · Posted 2015-04-21
Advertisement
Recalling FirmCaptiva Spine, Inc
Recall StatusTerminated
Product CodeLXH
Date Initiated2015-02-25
Date Terminated2015-06-01
Product Quantity233 devices
Reason for Recall
It is possible for the titanium color-coded ring to detach from the Reamers.
Firm Action
On 02/26/2015 the firm sent a Voluntary Correction Notification to its consignees via FedEx. Within that notification the firm asks to please check with your product representative and Central Supply Departments to determine if you have any of the products listed on the notificaiton. If any product is located please call 877-772-5571 from 8:30am (Eastern Time) to receive instructions as to how to remove the color-coded ring OR please follow the enclosed steps for the correction in their entirety. They contain information about returning the removed rings. Enclosed is Form 30.4 Product Tracking and Verification Form. You must complete and return this form even if you do not have any rings to remove and return. Please contact your Captiva Spine customer service representative if you have any questions regarding this correction, any of our products, or would like assistance with the correction.
Root Cause
Device Design
Distribution Pattern
GA, IN, FL, UT, AZ, CA, and NY
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.