Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA/DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test.
FDA Recall Z-1480-2018 · Posted None
Advertisement
Recalling FirmLife Technologies Corporation
Recall StatusTerminated
Product CodePQP
Date Initiated2018-02-08
Date Terminated2020-05-12
Product Quantity30
Reason for Recall
The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.
Firm Action
The firm issued an Urgent: Medical Device Recall Notification, dated February 16, 2018, via e-mail and phone call to affected consignees. The letter identified the affected device and advised customers of the potential "No Call" or false positive outcomes. Users were advised that they may need an alternate method to determine the presence of ROS1 variant in previously tested patient samples, and that any future tests using the affected lot cannot be considered valid if "No Call" or "Negative" is the result. Requested action included reviewing previous results which used the affected lot. Customers are to complete the Customer Response Sheet provided and e-mail a scanned copy along with the letter to [email protected].
Questions and comments should be directed to Vineet Bansal, 408-802-2341, [email protected].
Root Cause
Process control
Distribution Pattern
Distributed to states: AZ, CA, NC, TX, and VA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement