HSG Trays/HS 5F Catheter Combo; HSG Procedure Tray+ Shapeable 5F HSG catheter. For hysterosonography and hysterosalpingography, gynecological use.
FDA Recall Z-1481-2015 · Posted 2015-04-21
Advertisement
Recalling FirmCatheter Research, Inc
Recall StatusTerminated
Product CodeHGS
Date Initiated2015-04-07
Date Terminated2015-07-17
Product Quantity880 units
Reason for Recall
HSG Trays and Catheters were mislabeled. Trays with the flexible HS catheter were labeled as containing the shapeable, stiffer HS catheter, and vice versa.
Firm Action
Catheter Research, Inc. (CRI) has initiated a voluntary medical device removal of HSG Tray/ 5F HS Catheter Combo, and issued letters to customer son 04/07/2015 by e-mail and certified mail. Customers are asked to inspect their inventory and return recalled devices to CRI for replacement. If the devices were further distributed CRI requested customers provide notice of the recall. Customers may contact Babacar Diouf, VP of RA and QS, at (317) 872-0074 ext. 3512.
Root Cause
Packaging process control
Distribution Pattern
Domestic: MN, CA, OH, IL, FL, MA, WI, VA Foreign: None DOD/VA: None