LUMINOS Lotus Max (VF11)
FDA Recall Z-1481-2023 · Posted 2023-05-23
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2022-11-23
Product Quantity704 units in the United States (1945 units worldwide)
Reason for Recall
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
Firm Action
Siemens Medical Solutions USA, Inc., will provide software updates to the
affected systems with software versions VF10 and VF11 with software via Update
Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide - Worldwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.