Open, Classified

KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culture media

FDA Recall Z-1483-2024 · Posted 2024-04-05

Recalling FirmMicrobiologics Inc
Recall StatusOpen, Classified
Product CodeJTR
Date Initiated2024-03-06
Product Quantity19 units

Reason for Recall

The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when using the control material.

Firm Action

Microbiologics issued an Urgent Medical Device Recall notice to its consignees on 03/06/2024 via email. The notice explained problem with the device and provided the following instructions: 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage. 3. COMPLETE the response form provided. 4. RETURN the response form to [email protected] 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. For questions contact Recall Support team at 320.229.7080 or [email protected]

Root Cause

Use error

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, MI, KY, MO, NM, TN, PR, and the countries of France, Columbia, Kazakhstan, Peru, Puerto Rico, South Africa.