LVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.

FDA Recall Z-1484-2024 · Posted 2024-04-11

Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2024-03-07
Product Quantity17 units

Reason for Recall

Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.

Firm Action

A letter titled URGENT - Medical Device Field Correction, dated 3/7/24, was delivered by hand to consignees or by mail notifying them of this recall event. This notice details eleven software anomalies that could pose a potential risk of serious patient harm or death and the mitigations consignees should perform until software is updated. Consignees are to update their software to LVP software version 5.9.1 by reaching out to their Fresenius Kabi representative at 1-978-775-8100 to schedule installation. Consignees are to complete and return the Customer Reply Form via fax or email. Questions about the recall notice can be directed to Fresenius Kabi by email at [email protected] or by phone at 1-978-775-8100 (Monday through Friday from 8:30 am to 5:00 pm).

Root Cause

Software Design Change

Distribution Pattern

US Distribution: CA, CO, ID, MI, NE, NJ, NV, OK, TX, UT, VA, WA, & WI.