Merit Laureate Hydrophilic Guide Wire, Catalog No. LWSTFS35260EX, Straight Tip Stiff Shaft, 0.035" (0.89mm), 260 cm (102").
FDA Recall Z-1485-2014 · Posted 2014-04-22
Advertisement
Recalling FirmMerit Medical Systems, Inc.
Recall StatusTerminated
Product CodeDQX
Date Initiated2014-04-14
Date Terminated2014-07-17
Product Quantity173
Reason for Recall
Discrepancy between the carton and unit labeling for the Merit Laureate Hydrophilic 0.035" 260 cm Guide Wire, catalog number LWSTFS35260EX, which has a straight tip and a stiff shaft. This lot was partially mixed with another lot of Laureate
0.035" 260 cm Guide Wire, catalog number LWSTDA35260EX, which has an angled tip, standard shaft and Chinese unit labeling.
Firm Action
Customers were notified on 4/14/14 via certified letter with instructions to identify product, discontinue use, and return products.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution: USA Nationwide and China. No military/govt/VA consignees.