PureWick External Catheter for Women-Wick Product Usage: A flexible, contoured external catheter (a Wick) is positioned between the labia and gluteus muscles. The function of the Wick is to draw urine away from the body through the Wick to a storage container away from the body through a PVC tube via a continuous, low pressure vacuum pump (PureWick DryDoc) or to a hospital wall vacuum. There is never a reversal of flow direction or static state of fluid between the device and the patient. Also, the dry natural rubber component is double wrapped with 75D polyester and it is not stretched in its application to expose bare rubber to a user. The Wick does not enter the vagina (Figure 1).
FDA Recall Z-1485-2017 · Posted 2017-03-06
Advertisement
Recalling FirmPurewick Corporation
Recall StatusTerminated
Product CodeNNW
Date Initiated2016-02-24
Date Terminated2017-05-09
Product Quantity176,863 units
Reason for Recall
PureWick is recalling external catheter for women Wicks because the labeling did not contain the phrase "This product contains dry natural rubber" and/or erroneously included the phrase "Latex-free." The white gauze component of the PureWick contains dry natural rubber.
Firm Action
A Urgent Medical Device recall letter dated March 2, 2017 will be sent to customers to inform them that PureWick Corporation is recalling specific lots of PureWick external catheter for women Wicks, Item #625. Customers are informed that PureWick is recalling PureWick Wicks because labeling on product shipped prior to February 22, 2017 either did not contain the phrase This product contains dry natural rubber and/or erroneously included the phrase Latex-free. The white gauze component of the PureWick Wick contains dry natural rubber. PureWick Corporation has received zero complaints or reports of adverse events related to the error in labeling. Customers are informed of the risk to health and the actions to be taken. Customers are instructed to complete the enclosed PureWick Field Correction form and fax to PureWick Corporation at (619) 660-5459 or email to [email protected]. Customers with questions or would like assistance completing the PureWick Field Correction Form are instructed to call PureWick toll-free at (844) 584-0734 Monday Friday from 8:00 a.m. to 4:30 p.m., Pacific Time.
Root Cause
Labeling design
Distribution Pattern
US: Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.