Kuschall''s K3/K4 Series of Manual Wheelchair, Model Airlite.
FDA Recall Z-1486-06 · Posted 2006-09-06
Advertisement
Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeIOR
Date Initiated2006-07-18
Date Terminated2012-03-01
Product Quantity208 wheelchairs
Reason for Recall
If the user has chosen to install optional anti-tippers on the chair, the anti-tipper as designed may not be able to bear the stress of repeated or quick loads placed on it, causing the bolt to bend or break.
Firm Action
The recalling firm notified all consignees by recall letter sent on or about 7/20/06 along with replacement anti-tippers for each device on hand.
Root Cause
Device Design
Distribution Pattern
Nationwide and Canada.