ADVIA Centaur CA 125II Assay; Cat No. 09427226, SMN 10310443 (1-pack/100 tests); Cat no. 01678114, SMN 10315696 (5-pack/500 tests)

FDA Recall Z-1486-2018 · Posted None

Recall StatusTerminated
Product CodeLTK
Date Initiated2018-02-21
Date Terminated2019-05-16
Product QuantitySMN 10310443 (1-pack/100 tests) - 13257 kits; SMN 10315696 (5-pack/500 tests) - 9930 kits

Reason for Recall

Upon dilution, some patient samples may exhibit over-recovery outside the representative data provided in the assay instructions for use.

Firm Action

A Customer Notification was sent to customers on 2/21/18. The letter instructs customers to refer to the information provided until the appropriate updates are available in the assay instructions for use. The letter is to be retained and forwarded to those who may have received the product.

Root Cause

Other

Distribution Pattern

US Nationwide, and the following countries: Albania, Angola, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mexico, Morocco, Myamar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan,Thailand, Turkey, United Arab Emirates, Ukraine, United Kingdom, Uruguay, Uzbekistan,Vatican City, and Vietnam.