iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.
FDA Recall Z-1487-2014 · Posted 2014-04-22
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusTerminated
Product CodeIYN
Date Initiated2014-04-03
Date Terminated2015-03-18
Product Quantity9 units
Reason for Recall
Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
Firm Action
Boston Scientific sent an Urgent Medical Device Field Correction letter to all affected customers via hand delivery by BSC sales representatives. The letter identifed the affected product, problem and actions to be taken. Customers were asked to immediately post the Field Correction information in a visible location near the system(s) to ensure it is easily accessible to all users of the device. For question contact your local Sales Representative.
Root Cause
Nonconforming Material/Component
Distribution Pattern
USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.