VERIGENE Processor SP - Product Usage: intended for processing and identifying nucleic acids utilizing gold nanoparticle probe technology.
FDA Recall Z-1488-2021 · Posted None
Advertisement
Recalling FirmLuminex Corporation
Recall StatusOpen, Classified
Product CodePCH
Date Initiated2021-03-18
Product Quantity1210 Units
Reason for Recall
Potential for a false-negative result for Vibrio parahaemolyticus, Salmonella, and/or Shigella spp when used with the VERIGENE Enteric Pathogens Nucleic Acid Test Kit.
Firm Action
A customer advisory notice dated 03/18/2021. The Customer Advisory Notice recommends customers review their QC data to ensure passing results for all VERIGENE EP test targets, and to contact Luminex if any false negatives are observed for any targets within the past twelve months, in order to arrange for a replacement of the Amplification Heater. Additionally, if a Processor SP unit is indicated to have received service 15 on the TSB Log, the Amplification Heater would have already been replaced under routine maintenance with an in specification Amplification Heater and no further action is required for that instrument.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - United States (nationwide) and the countries of Austria, France, Greece, Japan, Netherlands, Turkey.