Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

dideco Preassembled Surgical Wash Set, Electa Essential , Cobe part numbers WS55E, WS125E, WS175E, WS225E, Cobe Cardiovascular, Inc.

FDA Recall Z-1489-06 · Posted 2006-09-09

Recall StatusTerminated
Product CodeCAC
Date Initiated2006-08-04
Date Terminated2011-12-28
Product Quantity8,964 wash sets, 1,518 Stat pacs-Recalls Z-1488-1491-06

Reason for Recall

Foreign particles-manufacturing error may lead to abrasion of bowl seal in blood recovery sets, possibly causing particles in the blood.

Firm Action

Consignees were notified by telephone and e-mail on 08/04/2006.

Root Cause

Other

Distribution Pattern

Nationwide, including VA facilities in CA and OH

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑