EG-3870UTK, Pentax Ultrasound Video Gastroscope. The EG-3870UTK, Ultrasound Video Gastroscope, is intended to provide optical visualization via a video monitor of, ultrasonic visualization of, and therapeutic access to the Upper Gastrointestinal Tract, including, but restricted to the organs, tissues, and sub-systems; Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the need for procedure are observed in adult and pediatric patient populations.

FDA Recall Z-1489-2012 · Posted 2012-05-03

Recall StatusTerminated
Product CodeITX
Date Initiated2012-03-14
Date Terminated2015-07-21
Product Quantity194 units

Reason for Recall

Endoscope ultrasound balloon evacuation tube may become clogged with patient body fluid if users fail to properly follow instructions for pre-procedure check, operating the suction valve during procedure and pre-cleaning procedure.

Firm Action

Pentax sent an "URGENT MEDICAL DEVICE RECALL" letter on March 14, 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. A response form was included for customers to complete and return to the firm. Contact the firm at 800-431-5880 ext. 2064 for questions regarding this notice.

Root Cause

Equipment maintenance

Distribution Pattern

Nationwide Distribution.