Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

CORDIS-SV-5 Steerable Guidewire, 300 cm.,5 Steerable Guidewires,Cordis a Johnson Johnson Company ,ENDOVASCULAR, Manufactured for: Cordis Corporation, Miami, FL 33102-5700. USA., Catalog # 503558X

FDA Recall Z-1493-06 · Posted 2006-09-09

Recalling FirmCordis Corporation
Recall StatusTerminated
Product CodeDXQ
Date Initiated2006-07-21
Date Terminated2008-11-19
Product Quantity480

Reason for Recall

Tip separation-Cordis SV-5 and SV-8 Steerable Guidewires may have a potential for guidewire fracture resulting in tip separation.

Firm Action

A recall of all lots manufactured by the OEM was initiated by letter dated July 21, 2006. All lots of the SV-5 guidewire and related design SV-8 guidewire were placed on Stop Shipment 6/30/06.

Root Cause

Other

Distribution Pattern

World wide Distribution-USA and Japan, Canada, Panama, Peru, Puerto Rico, Uruguay, Mexico and The Netherlands

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑