TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)

FDA Recall Z-1493-2019 · Posted None

Recall StatusTerminated
Product CodeOAE
Date Initiated2019-04-05
Date Terminated2021-04-27
Product Quantity317 units

Reason for Recall

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

Firm Action

The recall notice, and clinician acknowledgment form were hand delivered by Abbott Sales Representatives to all impacted accounts starting April 5, 2019. Completed Clinician Acknowledgement Forms, signify they received the correction notification for the impacted TactiSys Quartz Equipment with software version 1.7.

Root Cause

Software design

Distribution Pattern

US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA

Official Source

View FDA Recall →