Chromosome X Alpha-Satellite in FITC Spectrum - Product Usage: Analyte Specific Reagents. Model: MPA3481
FDA Recall Z-1493-2021 · Posted None
Advertisement
Recalling FirmCytocell Ltd.
Recall StatusOpen, Classified
Product CodeMVU
Date Initiated2021-03-19
Product Quantity89 probes
Reason for Recall
Users may observe additional locus specific signals at 9q34. The device pack insert indicates the probe has no known cross-reactivity to 9q34.
Firm Action
On 3/19/2021, Cytocell issued an Urgent Medical Device Recall notice via email notifying customers they may observe additional locus specific signals at 9q34. The device pack insert indicates the probe has no known cross-reactivity to 9q34.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of TX, OH, NY, IL, CT, AL, IA, MO.