Skytron General Purpose Surgical Table; Model 6600, Dist. by Skytron, Grand Rapids, MI.

FDA Recall Z-1496-06 · Posted 2006-09-13

Recall StatusTerminated
Product CodeFQO
Date Initiated2005-08-04
Date Terminated2008-05-01
Product Quantity160 total

Reason for Recall

The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.

Firm Action

A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables.

Root Cause

Other

Distribution Pattern

Nationwide.