CROSSCHECK(R) Plating system, Utility Plate, 4-Hole, REF CCP-UTN4. Orthopedic use.
FDA Recall Z-1496-2015 · Posted 2015-04-22
Advertisement
Recalling FirmWright Medical Technology, Inc.
Recall StatusTerminated
Product CodeHRS
Date Initiated2015-03-06
Date Terminated2018-08-05
Product Quantity372 units
Reason for Recall
The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction
Firm Action
Firm has initiated the recall on 03/06/2015 through written communication via Fed Ex P1 delivery to all Wright Medical Technologies distributors (hubs). Follow-up to non-responding consignees will commence 30 days after initiation. Effectiveness checks will be performed through the confirmation of returned product via a validated ERP system, a returned/signed verification form, and in the case of non-responders, a phone call, e-mail or follow up letter, depending on the region and availability of consignee.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.