Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Urine Combo 55; Catalog Number B1017-409. Siemens Healthcare Diagnostics, West Sacramento, CA.

FDA Recall Z-1497-2012 · Posted 2012-05-08

Recall StatusTerminated
Product CodeJWY
Date Initiated2011-06-16
Date Terminated2012-05-08
Product Quantity687 boxes (20 panels per box).

Reason for Recall

Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels.

Firm Action

Customer Notification letters were sent to all customers of the three MicroScan panels by traceable mail. The letter identified the affected product and asked customers to discard any remaining inventory in addition to contacting customer service for replacement product. The letter also recommended that customers consider the need to review previous test results, conduct patient follow-up, and or repeat tests as needed. Customers were to complete and return the customer reply card to confirm that they received and understood the recall communication.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- US, Turkey, Poland, and Great Britain.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑