7F Prelude Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.
FDA Recall Z-1497-2017 · Posted 2017-03-24
Advertisement
Recalling FirmMerit Medical Systems, Inc.
Recall StatusTerminated
Product CodeDYB
Date Initiated2017-02-16
Date Terminated2017-08-08
Product Quantity1,265 total
Reason for Recall
Merit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude Short Sheath Introducers because the sheath marker tips may detach during use.
Firm Action
Consignees notified on 2/16/17 to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact your Merit Sales Representative or Merit Customer Service at 801.208.4381.
Root Cause
Process control
Distribution Pattern
Nationwide and other foreign consignees. No Canadian or VA/govt/military consignees.