Recall StatusOpen, Classified
Date Initiated2025-03-12
Product Quantity293660 units
Reason for Recall
The potential for cracks forming in the breathing circuit hose.
Firm Action
On March 12, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken
When using these breathing circuits:
1. Use holders with a larger contact surface (see Figure 2 above).
2. Perform leak testing after stretching the flexible hose to the required length for the application. Please
observe the products Instructions for Use (IFU)
According to our records, you have received at least one of the listed breathing circuits. Please inform all potential
users in your facility. If you have forwarded affected products to third parties, please also forward this Urgent Medical
Device Recall notification. Please also complete the enclosed Medical Device Recall Return Response
Acknowledgment and Receipt Form to confirm to us that you have received this information.
Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch
Adverse Event Reporting program either online or by phone at 1-800-FDA-1088.
We apologize for any inconveniences resulting from this action. If you have any questions regarding this letter,
please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at
[email protected].
The US Food and Drug Administration has been advised of this action.
****Update 07/08/2025****On June 25, 2025 URGENT MEDICAL DEVICE RECALL - UPDATE letters were sent to consignees to clarify and provide complete UDI information for the device to include both primary and secondary UDI numbers as errors were identified in the initial letter.
Root Cause
Under Investigation by firm
Distribution Pattern
Domestic: Nationwide Distribution.
International - following countries:
Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, France, Germany, Ghana, Greece, Guatemala, Hungary, India, Italy, Japan, Panama, Philippines, Portugal, Singapore, South Africa, Spain, Sweden, United Kingdom