MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Breakpoint Combo Type 34, catalog number: B1017-404. Siemens Healthcare Diagnostics, West Sacramento, CA.
FDA Recall Z-1498-2012 · Posted 2012-05-08
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJWY
Date Initiated2011-06-16
Date Terminated2012-05-08
Product Quantity680 boxes (20 panels per box)
Reason for Recall
Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels.
Firm Action
Customer Notification letters were sent to all customers of the three MicroScan panels by traceable mail. The letter identified the affected product and asked customers to discard any remaining inventory in addition to contacting customer service for replacement product. The letter also recommended that customers consider the need to review previous test results, conduct patient follow-up, and or repeat tests as needed. Customers were to complete and return the customer reply card to confirm that they received and understood the recall communication.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution -- US, Turkey, Poland, and Great Britain.