Recall StatusTerminated
Date Initiated2020-02-24
Date Terminated2022-08-04
Product QuantityTotal Worldwide=9309 systems (2713 system in U.S.)
Reason for Recall
The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
Firm Action
On 02/24/2020, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers informing them that the firm is executing a voluntary correction for their Heater-Cooler Systems and providing customers with a recently cleared operating instructions that includes a validated instructions for cleaning and disinfecting the external surfaces, connectors, fittings, and water circuits. The firm is instructing customers to discard or destroy older versions of the operating instruction and advising that all Heater-Cooler Systems in use at their facility should be upgraded with the 3T Aerosol Collection Set (vacuum canister and internal seal) to further reduce the risk of potential aerosol emission form the 3T system.
These corrections apply to all 3T Systems currently in distribution, which includes upgrading any loaners that were previously distributed or that are pending distribution.
Actions to be Taken by the Customer/User
Use the Validated Operating Instructions Provided with this Customer Letter
" The FDA cleared Operating Instructions are found in Attachment 1 as well as at Product Resources on www.livanova.com.
- Please read and follow the instructions carefully.
- Section 2.1.5 Integrating the HC3T into your facility provides guidance on the use of the 3T System in your facility, highlighting critical tasks that users had the most difficulty with during human factors testing.
- Section 6 Maintaining the heater-cooler provides guidance on the maintenance of the 3T System including the validated cleaning and disinfection instructions.
- Section 7.1.4 General performance data describes performance testing results that emphasize the importance of following the cleaning and disinfection instructions and consistently using the 3T Aerosol Collection Set during surgeries. Specifically, these results describe how (1) the disinfection and water maintenance process maintains water quality and (2) the 3T Aerosol Collection Set design change reduces the
Distribution Pattern
U.S. Nationwide distribution including Puerto Rico.
This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected.
O.U.S. - none