ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
FDA Recall Z-1498-2023 · Posted 2023-05-02
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2023-02-02
Product Quantity141 units US
Reason for Recall
If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen
Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Roadmap workflow does not match real anatomy). This may cause e.g., a vessel perforation in DSA roadmap.
Firm Action
Siemens Healthineers issued to all users of ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM) a Customer Safety Advisory Notice to affected customers via AX041/22/S dated 2/3/23. Letter states reason for recall, health risk and action to take for addressing three potential software issues: Detection of wrong table movement direction; Adjust C-arm to Ref; Missing movement stop with multiple collision sensor activation. Siemens will correct the software error via Update Instructions AX040/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice.
Siemens
A software update to correct the error will be distributed to all affected customers via Update Instructions AX040/22/S. Affected systems will be updated to SW version VE10B.
Root Cause
Software design
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.