EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses (IOLs).

FDA Recall Z-1500-2017 · Posted None

Recalling FirmHaag-Streit USA Inc
Recall StatusTerminated
Product CodeNFJ
Date Initiated2017-01-05
Date Terminated2019-03-22
Product Quantity163 units

Reason for Recall

There is a possibility for data to be stored under the wrong patient on the DICOM Server following a non-standard workflow.

Firm Action

On January 5, 2017 the firm sent Field Safety Notice to their customers.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.