Neoligaments Staple Impactor, 8 mm Staple Impactor: 202-3010

FDA Recall Z-1501-06 · Posted 2006-09-13

Recalling FirmXiros Plc
Recall StatusTerminated
Product CodeHXJ
Date Initiated2006-05-17
Date Terminated2012-09-14
Product Quantity10, none distributed in US

Reason for Recall

The impactors are being recalled for modification so that they can be disassembled fully for cleaning and sterilization, on suspicion that the assembled items retain contamination after cleaning and may not be effectively sterilized by the recommended hospital sterilization cycles.

Firm Action

Customer recall notifications included: a) Telephone notification to highest volume users b) Recall letters were sent via certified mail to all customers June 1, 2006. Customers were instructed to cease using immediately and return the Impactors to Tapco. The customers were instructed to call the firm for a return authorization number. Modified loaner Impactors will be made available to customers as soon as possible.

Root Cause

Other

Distribution Pattern

Nationwide & Worldwide, including Spain, Japan, UAE, Italy, Germany