Terminated

Sickle- Scan Hemoglobin S Cell Reagent Set, Catalog No. 13-2101 100 Tests For the Determination of Hemoglobin S Two bottles Sickle Cell Solubility Buffer, 100 ml , 2 vials Sickle Cell Solubility Powder, Size 100 tests, Store 20-30 C, Distributed by True-Medix Diagnostic Brooklyn, NY 11228, 800-952-3349. For the qualitative determination Hemoglobin S, for in vitro diagnostic use only.

FDA Recall Z-1503-2012 · Posted 2012-05-09

Recall StatusTerminated
Product CodeGHM
Date Initiated2012-02-23
Date Terminated2012-12-13
Product Quantity400 kits

Reason for Recall

Two complaints concerning the powder vial included in the 100 test kit has been observed producing pressure while sealed. When the powders are opened, the pressure from within the vial releases a small amount of powder into the air. This powder. if released near the face and inhaled can cause minor irritation resulting in sneezing and nasal congestion. Investigation found no affect on the perf

Firm Action

Michclone Associates, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 15, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to identify affected lots, notify sub accounts, and complete and return the response form to arrange return of unused product and replacement . Forms can be returned by FAX 248(307)-9498, mail: Michclone Associates Inc. 680 Ajax Drive, Madison Heights, MI 48118 or Email: [email protected]. Questions and concerns should be addressed to Technical Support Department at 248-307-9494.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →