MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.

FDA Recall Z-1505-2014 · Posted 2014-04-24

Recall StatusTerminated
Product CodeKPR
Date Initiated2014-02-13
Date Terminated2015-03-16
Product Quantity12

Reason for Recall

It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b

Firm Action

Affected customers were initially notified with a recall letter on/about 2/12/14 and were sent Update Instructions to affected customers.

Root Cause

Process change control

Distribution Pattern

Distributed in the states of AK, PA, KY, OH, WI, MI, FL, and NY.