Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Scooter

FDA Recall Z-1505-2021 · Posted None

Recall StatusOpen, Classified
Product CodeINI
Date Initiated2021-04-01
Product Quantity189 units

Reason for Recall

Seat baseplate pin has broken in the field.

Firm Action

A Recall notification letter titled, "MEDICAL DEVICE FIELD CORRECTION" was sent to consignees on 04/07/2021 via email and by certified mail on 04/16/2021. The firm stated they made phone call notifications on 04/01/2021. The letter instructs the consignee to Contact impacted customer using a personalized customer letter (template provided). The notification can be sent by mail, email or phone. (IF by phone, please keep a log of call. Date/Time/Result); replace the old seat plate with new one provided using the SOP provided. The field correction should take approximately 15 minutes to complete.; complete and return the Service and Repair Report (sent with new part), inspect the installation and note it on the appropriate Scooter or Power Chair Inspection Record provided; use the RA shipping label (sent with new part) and return old seat plate to Merits Health Products along with the Service and Repair Report (one filled out for each unit); and when all of the units in the possession of the consignee are finished, please return the completed Inspection Report/s to Merits by mail or email.

Root Cause

Process change control

Distribution Pattern

Distributed nationwide to AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MN, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WV, WI and internationally to Canada

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑