ADVIA Centaur Cortisol 50T (Material Number 10994924), ADVIA Centaur Cortisol 250T (Material Number 10994926), and ADVIA Centaur Cortisol REF 250T (Material Number 10994927). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
FDA Recall Z-1505-2023 · Posted 2023-05-03
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusOpen, Classified
Product CodeJFT
Date Initiated2023-03-16
Product Quantity18,961 kits
Reason for Recall
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
Firm Action
An Urgent Medical Device Correction notification letter dated 3/17/23 was sent to customers.
Actions to be Taken by the Customer
Discontinue use of urine patient samples with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays until further notice.
Customers may continue to use serum and plasma sample types with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays.
Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable.
Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days.
If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Please retain this letter with your laboratory records and forward this letter to those who may have received this product.
We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
An additional Resolution Customer Notification letter dated 2/29/24 was sent to customers explaining an improvement to raw material specification that restores assay performance with urine samples. Customers are asked to follow the instructions for urine patient samples and urine cortisol QC as stated in the notification and if using products listed in the original field action notice continue following the guidance as stated in the Field Action Letter. Customers with any questions are to contact the Customer Care Center or local Siemens Healthineers technical support representative.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution - US Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.