Siemens Axiom Sensis XP diagnostic programmable computer Diagnostic programmable computer

FDA Recall Z-1506-2012 · Posted 2012-05-10

Recall StatusTerminated
Product CodeDQK
Date Initiated2012-02-22
Date Terminated2014-01-24
Product Quantity3

Reason for Recall

There is a potential malfunction in the connection of a Sensis EP system with Carto(R) 3 dedicated cable set (part number 14411343) with the Carto 3 system from Bionsense Webster.

Firm Action

Siemens sent a recall/correction letter dated February 22, 2012, via Update Instruction AX049/11/S of this product to all affected customers. The letter informed customers of the potential issues and provides additional information on the implemented resolution. An Update Instuction AX048/11/S will be available immediately. The update instruction provides a new version of the Carto 3 addendum to the Operator Manual and the exchange of the previous stim cables with the latest cable. For questions call 610-219-4834.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution including FL, NC, and SC.