Product is labeled in part: Inner Label "***SMART Control Transhepatic Biliary Stent***REF Cat. No.***Lot No.***Use By***Cordis S.M.A.R.T. CONTROL Nitinol Stent Transhepatic Biliary System***Assembled in Mexico***Cordis Corporation 14201 North West 60th Avenue Miami Lakes, Florida 33014, USA***Cordis S.M.A.R.T. CONTROL BILIARY***Cordis a Johnson & Johnson company***Endovascular***" Catalog C10080MB Lot No. 15535360. Catalog C10080SB Lot No. 15530747. Catalog C12030SB Lot No. 15530756. Catalog C12040SB Lot No. 15530755. Catalog C12060SB Lot No's. 15535363, 15569685. Catalog C12080MB Lot No. 15530757 Catalog C14040SB Lot No's. 15530754, 15557313. Catalog C14060SB Lot No's. 15530753, 15557311. Catalog C14080SB Lot No. 15535361. Indicated for palliation of malignant neoplasms in the biliary tree.
FDA Recall Z-1508-2012 · Posted 2012-05-10
Advertisement
Recalling FirmCordis Corporation
Recall StatusTerminated
Product CodeFGE
Date Initiated2012-04-09
Date Terminated2013-12-24
Product Quantity278 units
Reason for Recall
Cordis Corporation is recalling 32 lots of Cordis S.M.A.R.T. Nitinol Stent Transhepatic Biliary System and S.M.A.R.T. Control Nitinol Stent Transhepatic Biliary System. There is a potential for a sterility breach due to holes or tears in the Tyvek packaging.
Firm Action
Cordis Corp sent a " Urgent Medical Device Recall (Removal) letter dated April 9, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A Response Form was provided with each notification. Customers were asked to complete and return Response Form.
The customers were instructed to :
" Immediately identify and set aside all product listed above in a manner that
ensures the affected product will not be used.
" Review, complete, sign and return the enclosed Acknowledgement Form
directly to Cordis at the fax number on the form.
" Either return any affected product with reference to the RA# on the form, or
contact your local sales representative, or contact Customer Service (see
Available Assistance section) to facilitate return of the affected product.
Replacement product will be provided in a short time frame. Credit will be
provided for customer-owned inventory upon request.
" Pass on this notice to anyone in your facility that needs to be informed.
" If any product listed below has been forwarded to another facility, contact that
facility to arrange return.
" Maintain awareness of this notice until all product listed above has been
returned to Cordis.
For questions related to the issue or product replacement, please contact
your local sales representative, or contact the special Customer Service
number (800) 551-7683, Monday through Friday from 8:30 AM to 8:00 PM
EDT.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution-- USA (nationwide) and the country Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K062798SMART NITINOL SELF-EXPANDING STENT, MODEL C06120MB, C07120MB, C08120MB, C06150MB, C07150MB, C08150MBCordis Corp. · 2008-02-06K042969MODIFICATION TO SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEMCordis Corp. · 2004-11-08K031777MODEIFICATION TO: SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEMCordis Corp. · 2003-07-07