Axis Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143

FDA Recall Z-1510-2008 · Posted 2008-09-19

Recall StatusTerminated
Product CodeKPS
Date Initiated2007-11-21
Date Terminated2012-02-25
Product QuantityUnknow at this time since the recall was expanded.

Reason for Recall

Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.

Firm Action

Consignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Netherlands, Norway, Philipines, Poland, Singapore, South Africa, Spain, Switzerland, and United Kingdom.