cobas e801 Immunoassay Analyzer

FDA Recall Z-1515-2022 · Posted None

Recall StatusOpen, Classified
Product CodeJJE
Date Initiated2022-07-14
Product Quantity965 units

Reason for Recall

Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.

Firm Action

The firm distributed an Urgent Medical Device Correction notification via mail by UPS on 07/14/2022. The notice indicates that cobas e801 (Catalog No. 08454345001) and cobas e801 module (Catalog No. 04682913001) are being recalled due to a software issue affecting all analyzers, which may result in impacted signals and sample results of Pre-Wash tests when using Quick Start Mode for the following assays: Anti-HAV 2, Anti-HBc IgM, IGF-1, Myoglobin, Rubella IgG, and Toxo IgG. Customers with affected units are provided workaround instructions to deactivate Quick Start Mode or additional software maintenance with Quick Start Mode still activated. The firm states that the issue will be fixed in the next software version release. Customers are asked to return the attached Faxback form. Any questions are to be directed to Roche Customer Service at 1-800-428-2336.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, DC & PR.