ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.
FDA Recall Z-1516-06 · Posted 2006-09-23
Advertisement
Recalling FirmSiemens Healthcare Diagnostics
Recall StatusTerminated
Product CodeJJE
Date Initiated2006-08-04
Date Terminated2012-07-30
Product QuantityDomestic - 101, Foreign - 153
Reason for Recall
Sample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank
Firm Action
On 8/04/06, a Support Bulletin was e-mailed to all affected Bayer HealthCare LLC Branches worldwide in order to inform them of the issue and to provide them with instructions for this 'field correction' until the new software version is issued. A Customer Bulletin and Confirmation Form (FaxBack) were prepared and sent to the affected Bayer Healthcare LLC Branches on 8/04/06 for communication with the affected customers worldwide.
Root Cause
Software design
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.