GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

FDA Recall Z-1516-2008 · Posted 2008-09-29

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeMQB
Date Initiated2007-10-21
Date Terminated2011-12-17
Product Quantity146 total

Reason for Recall

Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c

Firm Action

GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.