IMMULITE /IMMULITE 1000 Systems TIE Total IgE, REF LKIE1, SMN 10380867, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.

FDA Recall Z-1517-2014 · Posted 2014-04-25

Recall StatusTerminated
Product CodeJHR
Date Initiated2013-12-20
Date Terminated2018-05-18
Product Quantity5953 units

Reason for Recall

Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.

Firm Action

Urgent Medical Device Correction Letters (dated 12/20/13) and effectiveness check response forms were sent to customers on 12/20/13 via FedEx to inform them of the issue. Customers were requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within thirty (30) days. Field service personnel were sent a support document describing the issue and instructing them how to deal with customer questions.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution.