Terminated

LICOX CMP Tissue Oxygen Monitor; Catalog numbers REF AC31. Manufacturer GMS Gesellschaft fur medizinische, Sondentechnik mbH, Dorfstrasse 2, 24247 Mielkendorf, Germany; Distributed by Integra NeuroSciences, 311 Enterprise Drive, Plainsboro, New Jersey 08536 USA. Used for continuous determination of oxygen partial pressure in cerebrospinal fluid and brain tissue.

FDA Recall Z-1518-2011 · Posted 2011-03-03

Recall StatusTerminated
Product CodeGWM
Date Initiated2010-12-17
Date Terminated2012-11-05
Product Quantity26 units

Reason for Recall

Certain LICOX CMP Tissue Oxygen Monitors could allow for underestimation of temperature values.

Firm Action

Integra Life Sciences notified their Integra Neurospecialist via email and their customers via Fed Ex overnight mail on December 17, 2010 with an Urgent: Product Recall letter. The letter identified the affected product and the reason for the recall. The letter states that an Integra Neurospecialist will contact customers to arrange for the return of their Licox monitor to the Service Center for recalibration. A loaner will be provided until the recalibration is complete. The attached "Recall Acknowledgement and Return Form" must be completed and returned to Integra. If there are any additional questions, customers should contact Gary Rabinovich at 609-936-6864.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution