Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test
FDA Recall Z-1519-2018 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Corporation
Recall StatusTerminated
Product CodeLDP
Date Initiated2018-01-02
Date Terminated2019-04-25
Product Quantity35,111
Reason for Recall
The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400
plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following
statement in the Limitations-interference section:
"In very rare cases, gammopathy, in particular type IgM (Waldenstrm s macroglobulinemia), may cause
unreliable results."
Firm Action
On January 2, 2018 an URGENT MEDICAL DEVICE CORRECTION letter was issued to customers with notification to the issue and information regarding the updated method sheets and that they were available for use. The COBAS INTEGRA 400 plus/800/800 CTS was updated from affected method sheet version 3.0 to 4.0. The MODULAR ANALYTICS P was updated from affected method sheet version 4.0 to 5.0 No product return is required. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week
at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Root Cause
Error in labeling
Distribution Pattern
Nationwide Distribution